- High definition colour touch-screen
- Simple screen icon interface using stylus or optional MicroMouse
- Fast on-screen textual entry enabling comments and report writing
- Fast, quiet printer with easy load paper mechanism
- Fully customisable printout format
- Direct connection to external Hewlett Packard printers.
- Unique contextual Help Screens (language selectable) with dedicated Help Button
- 41 inspiratory and expiratory indices measurable
- On-screen graphical child incentive device
- Large 2000 patient test memory
- Internationally recognised on-screen test quality assurance prompts (e.g. slow start, abrupt end etc.) ensuring quality test results
- Unique ‘today’s testing feature’ - creating a list of patients tested or to be tested
- Numerous predicted values and languages selectable
- ‘Lung Age’ assessment and textual result interpretation by ATS (9), BTS-NICE (10) or Enright (11)
- Body Mass Index and Dyspnoea Score
- Bronchodilator and/or Steroid assessment
- Complete with all accessories in a sturdy carrying case
|
|
| |
Spirometry |
| Measurements (forced): |
VC, FEV.75, FEV1, FEV3, FEV6, FVC, PEF, FEV.75/VC, FEV.75/FVC, FEV1/VC, FEV1/FVC (FER), FEV3/VC, FEV3/FVC, FEV.75/FEV6, FEV1/FEV6, FEF25 (MEF75), FEF50 (MEF50), FEF75 (MEF25), FEF25-75 (MMEF), FEF50/VC, FEF50/FVC, MMEF/FVC (FEF25-75/FVC), FIV1, FIVC, PIF, FIV1/FIVC (FIR), FIF25 (MIF75), FIF50 (MIF50), FIF75 (MIF25), R50 (FEF50/FIF50), MET25-75, FET, MVV (ind) |
| Measurements (relaxed): |
EVC, IVC, IC,VT (TV), ,Ti,Te,Ti/Ttot.,VT/Ti (TV/Ti), IRV, ERV, FR |
| Tests per subject: |
5 relaxed VC maneuvers and 8 forced maneuvers for each baseline and two post examinations |
| Predicted Values: |
Knudson, Crapo or NHANES III |
| Transducer: |
Micro Direct Gold Standard Bi-Directional Digital Volume |
| Resolution: |
10ml volume 0.03l/s flow |
| Accuracy: |
+/-3%.To ATS recommendations – Standardisation of spirometry 1994 update for flows and volumes |
| |
General |
| Storage: |
2000 patients’ tests including Flow/Volume loops and Volume/Time graphs |
Printer Output: (external printers) |
Compatible with a wide range of Hewlett Packard printers; contact us for details. |
Printer Output: (internal printer) |
0.5" per second (avg) print speed. |
| Power Supply: |
100 – 240V AC 50 – 60Hz. DC output 12V 2.5A |
| Battery Pack: |
Rechargeable NiMH 8.4V 1Ah |
| Dimensions: |
5" x 10" x 1.5".Transducer 2"x 2"x 3.5" |
| Weight: |
Excluding any transducers: 1 lb 7 oz. |
| Temperature: |
The instrument will operate in a uniform environment of 32°F - 96°F, out of direct sunlight |
| Operating Humidity: |
30% to 90% non-condensing |
| Storage Temperature: |
0 to +144°F |
| Storage Humidity: |
10% to 90% RH |
| Connectivity: |
RS232 serial and USB 1.1 | |
|